GE MR — field troubleshooting
9,114 field-service issues · page 72 of 114
Symptom, error message and equipment context for each GE MR issue are public — the full cause analysis and step-by-step resolution are available to FSELIB subscribers.
- Unable to reset TPS with AP Initialization errors Due to AP Board.MR · 2227556, 2227996
- Coil cable is damaged and exposing wires.MR
- MRIQ Shading in Images Due to the Coil.MR
- Intermittent Warnings of high Patient Comfort Temperature Readings due to Gradient Chiller.MR · 2248650
- ICN configuration fails with DHCP request due to defective SUN ICN.MR
- Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Scan Room Power Supply.MR · 2269058, 2269056
- Part Number for 8NV ARRAY Coil Latches (MRID).MR
- Fails TPS Reset when booting up to Applications or when Resetting TPS due to Guided Install Does Not Reflect the Correct Spectro Amp State After Upgrade to 20.1_IB1 Software due to a Software Problem.MR · 2267411, 29917
- Printer pausing due to camera.MR · 14140
- Environmental Locks activate due to restricted airflow for GOC.MR · 2269034
- Oil Leak Under Patient Table due to Hydraulic Fluid Spill Inside Table.MR
- Magnet Monitor 3 was not placed into Service Mode prior to Magnet or System Service.lcc300i2 alarm I2 - HePressure Cable Open.Use Magnet Monitor 3 Service Mode When Performing Service That Will Genera…MR
- The Download Of A Prescribed Series Failed due to 2nd RECEIVER II Board in RRF Chassis.MR · 2248010, 191321
- Images are in Autoview Window but not in Browser.MR
- Patient Alarm squeeze ball is not working Due to Worn Squeeze Ball.MR
- Unable to prescribe new exam, all scan Rx functions greyed out, with error "The lock exam path is null", due to image database software corrupted when operator was viewing complex images and prescribi…MR · 2224366
- 2267910 Power cabinet reporting 48V overvoltage due to PS1 48 V supply in PDU.MR · 2267600, 2267910
- Anterior section of coil has bent pins in coil plug due to misalignment during connection.MR
- AP Errors Due to PPC Module on AP board.MR · 2232902
- Commands to use on the Magnet MonitorMR
- Defective LCD monitor.MR
- SHeM cannot view Current, Diagnose or Current page on BroadbandMR
- Intermittent Ghosting or Smearing of Image Due to loose coil cable.MR
- ID192543. TPS reset fails due to P1 or P2 port was not connected properly to LPCA .MR · 2276804
- Fails TPS Reset when booting up to Applications or when Resetting TPS due to incorrect cabling to RFCB during installation.MR · 2276804, 2274203
- Error 2268844 due to defective IBC sensor.MR · 2268844, 2224666
- Cradle is making a rhythmic clunking noise do to broken wheel.MR
- Run 830 HFO cable kitMR
- RF Hub RFCB and PDB temperature sensor errors due to RFCB.MR · 2273806, 2273807
- Can not load 16 channel option key due to previously installed 8 channel option key.MR
- Fails TPS Reset when booting up to Applications or when Resetting TPS due to 48V Power Supply 1000W.MR · 2267728, 2269060
- Which files caues alarm to be sentMR
- Does Toucholc.pl send hour dataMR
- System Stops Scanning with Error 2275652 RF Hub detected a fault while scanning due to Fiber Optic Cable Run 2025-D Coiled, Bent, or Kinked into a Diameter Less Than 12 Inches, 30.5 cm.MR · 2275652
- YMS CD drive compatibility.MR
- 2274021 Coil Fault Short Circuit A multicoil bias fault was detected due to Coil cable.MR · 2224666, 2274021
- TPS download failed due to broadcast pings from AGP and SCP was turned off with.MR · 2223880, 2223881
- ID192694. The Home Flag on the Longitudinal Drive Belt was not passing through the Home Flag Detect Optical Coupler.MR
- Table Drifts Down when Docked due to Down Pedal Dock Spring.MR
- System keeps losing landmark location during the scan due to patient positioning pad being stuck under the cradle or carriage assembly.MR · 2268835, 2633351
- SPT fails to run the body section of the test Due to The Table will not advance to the 1135 locationMR
- Documentation is Wrong.MR · 51828221, 5670005
- 15.0 M4B IServ contains only PC version of Online Help DVD, but not the LINUX version.MR
- Magnacoustics Music System Documentation Missing or Need Tech Help Parts.MR
- RRF Calibration Error in RRF Data Path Diagnostics Due To UTNS Universal Transient Noise Suppressor.MR · 2247340
- Table is Sticking due to Cradle Release Block is Extended.Table Is Sticking and Unlatching When Moving In The Bore.Perform Cradle Release Block Retraction Adjustment-wqr.Optima MR450WMR
- Will Not Scan With RF Errors due to SRFD2.Will Not Scan With RF Errors.SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 wqr.Signa HDxT 15.0 2254208 - SRFD2 RF Amplifier High Voltage LowMR · 2254208
- Table Movement is Rough Due to Longitudinal drive belt was loose.MR
- Helium Gas Lines For Cold head Compressor Are Too Short.Helium Gas Line Extensions Can Be Ordered. - WQR.Magnet LCC.2250631 Return Extension Line.2250629 Supply Extension Line.MR
- Images Retrieved or Pulled from PACS to Scanner Look Like Screen Saves instead of Normal Images due to Third Party PACS.MR
- MRSPT Fails FSE Stability Phase Offset and Ghosting Black Banding in Images due to Helium Line Routing.MR
- What is the Coil Config for the Medical Advances Shoulder Coil 415GE-64AMR
- Will Not Scan With RF Errors Due To Power Supply for ASC and DCERD.MR · 2254055, 2248847
- 2252310 DD Driver Fault Detected - Open Circuit Due to RF Body Coil.MR · 2252310
- Unable to Scan with Driver Module Errors due to Driver Module.MR · 2252312, 2252293
- GP Errors During DC Offset Calibration Due To GP3.MR · 2245067, 30000
- MRIQ Ghosting on all images in both phase and frequency directions due to new LED light bulb.MR
- Intermittent Power Monitor Faults in Head Amplitude during Body Transmit due to SRFD2.MR · 2254051, 2254052
- 2254781 Invalid power fault due to RF detector board in UPM.MR · 2224666, 2254781
- Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.MR · 2224842
- Z axis XPS module reports the loss of 420V power due to defective XPSMR · 2268224
- Cannot load Linux Operating System Software on HP PC due to DASM was Not Powered OffMR
- SHOCK SENSOR ON COIL IS RED Due to coil dropped.MR
- Unable to scan with error 2268843 or 2268844 due to gradient or RF noise during scanning.MR · 2268843
- iLinq failure due to corrupt iLinq software.MR
- 2254265 3T RF Amplifier PA Bias Current fault 61 due to RF amp.MR · 2254271, 2254265
- MRIQ Head Coil Images have area of Signal Loss Due to Coil Cable.MR
- Anatomy Outside the Field of View Wrapping Into Image, Even Though No Phase Wrap NPW Option Was Selected On, Due to No Phase Wrap By Design Only Corrects Up to Twice the FOV.MR
- Low signal on one side of the breast coil due to coil.MR
- How to Tell if a Coil is configured as Receive only or Transmit and Receive.MR
- Dock Pedal Stuck in Down Position Due to Dock Pedal Spring.MR
- Fails TPS Reset when booting up to Applications or when Resetting TPS due to Driver Module powered off.MR · 2252239, 1121430417
- ID 193081 Coil Not Recognized Due To Bent Pins In Quick Disconnect Box.MR · 2258107
- 2225540 Dynamic Disable Bias Fault on Front Drivers Due to the SSM.MR · 2225539, 2225540
- Can't Display Images Written to CDR or DVD on Excite HD using Centricity DICOM Viewer version 2.2 on a PC or Laptop due to Incorrect Jumper Setting of SCSI Bridge Board in SCSI Tower.MR · 35390
- The GOC network Hub was restricting the network speed.MR
- 2254250 MKS error code 35, 50V Power Supply Low Due to 3T MKS 35 Kw RF Amp.MR · 2254250
- 16 channel switch diag fail channel 13 on multiple input/outputs due to16 channel switch, importantly with RRF chassis system loopback diags passing. 16 Channel Switch Diagnostics Fail.MR · 2247386, 18571
- Operator Accidentally Hit Combination of Keyboard and Mouse That Caused Operator Graphical User Interface Window to Move.MR
- Current Distortion Faults Due to HFA CONTROL BOARD ASSEMBLY.MR · 2244578
Full cause & resolution for every GE MR issue