GE Healthcare · MRExciteDiscovery2224366
Unable to prescribe new exam, all scan Rx functions greyed out, with error "The lock exam path is null", due to image da…
Unable to prescribe new exam, all scan Rx functions greyed out, with error "The lock exam path is null", due to image database software corrupted when operator was viewing complex images and prescribi…
Symptom
Unable to prescribe new exam, all scan Rx functions greyed out.
Error message
2224366 The lock Exam Path is Null See Long Description Error: 2224366 Host: A091MRAGEM01 Process: TIR File: tirexec.c Line: 1949 The lock exam path is null
Applies to
- Models
- Excite, Discovery
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution