Pular para o conteúdo principal
GE Healthcare · MRDiscovery22690582269056226905422690592269057

Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Scan Room Power Supply.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2269058 The 1-Wire Network did not detect Clock Reference 12V Power 2269058 Task ID: 0x6db4420 Time: 4251 Object: AGPException Class: Notification Function: Software/Firmware/Config: N/A The 1-Wire Network did not detect Clock Reference 12V Power.->

2269056 The 1-Wire Network did not detect Exciter Digital Power. Error: 2269056 Host: andys3T Process: diagCoord File: oneWire.c Line: 1085 Task ID: 0x6db4420 Time: 4251 Object: AGPThe 1-Wire Network did not detect Exciter Digital Power.

2269054 The 1-Wire Network did not detect the Exciter 80MHz Clock that comes from the Clock Reference Board.

2269059 The 1-Wire Network did not detect the Clock Reference 80MHz Clock that is generated in the Exciter Module 2269059 Task ID: 0x6db4420 Time: 4251 Object: AGPException Class: Notification Function: Software/Firmware/Config: N/A The 1-Wire Network did not detect the Clock Reference 80MHz Clock that is generated in the Exciter Module.->

2269057 The 1-Wire Network did not detect light on the DVMR Fiber Optic Link between the Exciter and the IRF3 Board. Error: 2269057 Host: andys3T Process: diagCoord File: oneWire.c Line: 1098 Task ID: 0x6db4420 Time: 4251 Object: AGP The 1-Wire Network did 

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5231801-3 Scan Room Power Supply 16 ch and 32 ch 5231801-3 Scan Room Power Supply, 16 ch and 32 ch
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.