GE Healthcare · MRDiscovery225231222522932218537
Unable to Scan with Driver Module Errors due to Driver Module.
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252312 - DD Driver Fault Detected - High Voltage Fault. Error: 2252312 Host: mr3t Process: NSP File: SCP:DCB Line: 1884 DD Driver Fault Detected - High Voltage Fault. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value = 213 V Limit: 425 V-> 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.-> 2218537 Scan stopped, Scan Stopped due to error. Driver Module error occurred.. 2218537 HCMR scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : Driver Module error occurred..
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5160446-2 Driver Module NMR DVMR
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.