GE Healthcare · MRSigna2224842
Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.
Symptom
Cannot Advance Cradle to Scan Position See Long Description Cradle will not move into bore to landmark, longitudinal motor works but cradle is stuck on patient transport.
Error message
2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec-
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 46-282359P2 Gear 16 mm or 40-1 speed reducer Part of the longitudinal motor drive assembly in line with motor in rear pedestal. See long description 15 46–282359P2 2 SPEED REDUCE 1 40 TO 1 NON–MAGNETIC GEAR REDUCER Note: Part Description in FEMC is simply "GEAR 16 MM"
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.