GE Healthcare · MRDiscovery2276804
ID192543. TPS reset fails due to P1 or P2 port was not connected properly to LPCA .
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2 See Long Description for extended error. 2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2. Extended Error: The RF Hub detected that one of the coil ports in the LPCA was disconnected. Specific errors are logged in other messages. Re-seat the port P1 and P2 connector assemblies into the emergency egress connectors inside the LPCA. Check the coil port cable connections at the RF Hub control board and at the LPCA. Run the Coil Datapath Diagnostic that utilizes the MCRv tool to isolate the problem. Possible cause(s): RF Hub Control Board
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.