Pular para o conteúdo principal
GE Healthcare · MRSigna

Operator Accidentally Hit Combination of Keyboard and Mouse That Caused Operator Graphical User Interface Window to Move…

Operator Accidentally Hit Combination of Keyboard and Mouse That Caused Operator Graphical User Interface Window to Move.

Symptom

The Window which contains the Auto Prescan, Manual Prescan, Prep Scan and Scan buttons is shifted over to the right side and up. Can't scan because the window at the bottom of the screen which contains the "Auto Prescan", "Manual Prescan", "Prep Scan", and "Scan" buttons, was shifted over to the right side and up. It was partially obscured by the window above it and part of it was "hanging off the screen" on the right side, so that part of this window was invisible. Can't scan because the window at the bottom of the screen which contains the "Auto Prescan", "Manual Prescan", "Prep Scan", and "Scan" buttons, was shifted over to the right side and up. It was partially obscured by the window above it and part of it was "hanging off the screen" on the right side, so that part of this window was invisible.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution