System keeps losing landmark location during the scan due to patient positioning pad being stuck under the cradle or car…

System keeps losing landmark location during the scan due to patient positioning pad being stuck under the cradle or carriage assembly.

Symptom

System keeps losing landmark location during the scan

Error message

2268835 End of travel switch detected at incorrect location, must reset home location See Long Description for Extended Error message Error: 2268835 Host: BCMR02 Process: NSP File: SCP:LongMotionMonit Line: 2317 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: VIBRANT Coil: HD Breast Scan: 5056/9 End of travel switch detected at incorrect location, must reset home location. Expected Position=2633351 um Actual Position=2089846 Extended Error : This can be due to invalid scan range configuration in guided install or improper location of the end of travel switch on the bridge. If both of these are correct, the encoder may be malfunctioning and reporting incorrect position.

2218537 Scan stopped, Scan Stopped due to error The table is not at scan plane See Long Description 2218537 MsaScanMgr.m 2852 Scan stopped, Scan Stopped due to error : The table is not at scan plane..

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.