GE Healthcare · MRDiscovery2268835263335120898462218537
System keeps losing landmark location during the scan due to patient positioning pad being stuck under the cradle or car…
System keeps losing landmark location during the scan due to patient positioning pad being stuck under the cradle or carriage assembly.
Symptom
System keeps losing landmark location during the scan
Error message
2268835 End of travel switch detected at incorrect location, must reset home location See Long Description for Extended Error message Error: 2268835 Host: BCMR02 Process: NSP File: SCP:LongMotionMonit Line: 2317 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: VIBRANT Coil: HD Breast Scan: 5056/9 End of travel switch detected at incorrect location, must reset home location. Expected Position=2633351 um Actual Position=2089846 Extended Error : This can be due to invalid scan range configuration in guided install or improper location of the end of travel switch on the bridge. If both of these are correct, the encoder may be malfunctioning and reporting incorrect position. 2218537 Scan stopped, Scan Stopped due to error The table is not at scan plane See Long Description 2218537 MsaScanMgr.m 2852 Scan stopped, Scan Stopped due to error : The table is not at scan plane..
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.