GE Healthcare · MRSigna2269034

Environmental Locks activate due to restricted airflow for GOC.

Symptom

Environmental Locks activate when starting a new scan Environmental Locks activate when starting each scan. Operator is prompted to bypass the Environmental Locks before a new scan can be started.

Error message

2269034 The operator room temperature has exceeded the upper warning limits. Please make the necessary adjustments to prevent equipment damage. Wed Dec 29 09:17:09 2010 Error: 2269034 Host: DVMR Process: envMon File: monitor.c Line: 503 Exception Class: Abort Function: Global Operator Console: Environmental Sensors/Controls The operator room temperature has exceeded the maximum upper limits. Please make the necessary adjustments to prevent equipment damage. Instance: 0 Value instance: 0 Value: 35.31 Upper Limit: 35.00

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --