System Stops Scanning with Error 2275652 RF Hub detected a fault while scanning due to Fiber Optic Cable Run 2025-D Coil…

System Stops Scanning with Error 2275652 RF Hub detected a fault while scanning due to Fiber Optic Cable Run 2025-D Coiled, Bent, or Kinked into a Diameter Less Than 12 Inches, 30.5 cm.

Symptom

System Stops Scanning with Error 2275652 RF Hub detected a fault while scanning. See Long Description System scanning suddenly stops and only message reported in error log at time scan stopped is reported below: 2275652 ERROR: EM_RF_HUB_WAKEUP_STOP_SCAN RF Hub detected a fault while scanning.

Error message

2275652 RF Hub detected a fault while scanning See Long Description 2275652ERROR: EM_RF_HUB_WAKEUP_STOP_SCAN RF Hub detected a fault while scanning.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution