GE Healthcare · MRSigna2254250
2254250 MKS error code 35, 50V Power Supply Low Due to 3T MKS 35 Kw RF Amp.
Symptom
Unable to Scan with 3T MKS RF Amplifier will Not go Ready MKS amp powers up but will not come ready.
Error message
2254250 - 3T RF Amplifier 50 Volt Supply Low (fault 35). The SSD 50 Volt supply was below 5% specification. See Long Description for extended error message. 2254250 3T RF Amplifier 50 Volt Supply Low (fault 35). The SSD 50 Volt supply was below 5% specification. Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier. Extended Error: The RF Amp 50 V Power Supply has low voltage. Check fuses F15, F16 and F17, replace if necessary. If the fuses are functional, replace the power supply deck.
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT Apps For use with HD16.0_VO1 and HD16.0_VO2 version software only. For use with HD16.0_VO1 and HD16.0_VO2 version software only.
- Software
- HD16 Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only. Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only.
- Component
- SecondOpinion MKS 3T 35kW RF Amp 5104896 See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.