GE Healthcare · MRSigna2248010191321
The Download Of A Prescribed Series Failed due to 2nd RECEIVER II Board in RRF Chassis.
Symptom
The Download Of A Prescribed Series Failed
Error message
2248010 A/D Offsets is out of Valid Range See Long Description 2248010 File: DRF Line: 350 Diagnostic: RRF Calib Iteration: 1 of 1 Action: 1 Task ID: 0xa55688 Time: 191321 Object: APS A/D Offsets is out of valid range The calculated value is 1, valid range is 127,-128 (rcvr=0) Suspect analog receiver problem, please run Diagnostics!
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 5250094 RECEIVER 2A BOARD 5250094 RECEIVER 2A BOARD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.