GE Healthcare · MRSignaExcite

The GOC network Hub was restricting the network speed.

Symptom

Network Transmission Time is Slow. Takes 1/2 hour to transfer 1 series to the PACS. See long description The system is a 1.5T with Excite, running 11.0 M4 software. Network transfer is really slow. Sending an exam to the PACS and it took 1/2 hour for 1 series. We also see some packet loss when pinging the PACS (see results below). The FE connected his laptop to the hospital network port and was able to ping with no packet loss. It appears the problem is in the GOC. {sdc@VHALASMR2}[3] cd ~sdc/Prefs {sdc@VHALASMR2}[4] more SdCRHosts 10.176.12.13? vhalasnwg2 # 1 104 VHALASNWG2? PACS? 0 1 1 0 {sdc@VHALASMR2}[5] ping -s 1024 10.176.12.13 PING 10.176.12.13 (10.176.12.13) from 10.176.12.31 : 1024(1052) bytes of data. 1032 bytes from 10.176.12.13: icmp_seq=1 ttl=255 time=0.918 ms 1032 bytes from 10.176.12.13: icmp_seq=2 ttl=255 time=0.919 ms 1032 bytes from 10.176.12.13: icmp_seq=3 ttl=255 time=0.824 ms 1032 bytes from 10.176.12.13: icmp_seq=7 ttl=255 time=0.914 ms 1032 bytes from 10.176.12.13: icmp_seq=8 ttl=255 time=0.911 ms 1032 bytes from 10.176.12.13: icmp_seq=9 ttl=255 time=1.03 ms 1032 bytes from 10.176.12.13: icmp_seq=10 ttl=255 time=0.900 ms 1032 bytes from 10.176.12.13: icmp_seq=11 ttl=255 time=1.02 ms 1032 bytes from 10.176.12.13: icmp_seq=14 ttl=255 time=0.887 ms 1032 bytes from 10.176.12.13: icmp_seq=15 ttl=255 time=0.886 ms 1032 bytes from 10.176.12.13: icmp_seq=17 ttl=255 time=0.881 ms 1032 bytes from 10.176.12.13: icmp_seq=18 ttl=255 time=0.880 ms 1032 bytes from 10.176.12.13: icmp_seq=19 ttl=255 time=0.843 ms 1032 bytes from 10.176.12.13: icmp_seq=21 ttl=255 time=0.837 ms 1032 bytes from 10.176.12.13: icmp_seq=22 ttl=255 time=0.947 ms 1032 bytes from 10.176.12.13: icmp_seq=23 ttl=255 time=0.900 ms 1032 bytes from 10.176.12.13: icmp_seq=24 ttl=255 time=0.895 ms --- 10.176.12.13 ping statistics --- 25 packets transmitted, 17 received, 32% loss, time 24041ms <--- 32% packet loss! Round trip times look slightly higher rtt min/avg/max/mdev = 0.824/0.905/1.034/0.066 ms than expected as well.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5139480-2 8 PORT GIGABIT ETHERNET SWITCHING HUB See long description for important note FEMC Description: 8-Port Switch Kit with 4 cushions To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS. Note: BE CAREFUL when ordering this part. Per the FRU manuals, some products use this part for GOC and/or System Cabinet and some don't. Some products are supposed to use the 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT as the FRU. Please check the FRU manual for your system to see whether this part should really be used on the Product and for the purpose (i.e. GOC and/or Sys Cab).
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues