B2640000 — Error in XC
Canon Medical (Toshiba) CT error code, reported by the XC (X-ray Controller).
What B2640000 means
On Alexion and Activion16 and Aquilion RX CT systems, status code B2640000 indicates: Error in XC. The code originates in the X-ray controller (XC) / exposure section of the XC (X-ray Controller). Severity level A1 (Alarm — high severity).
Reported action: Confirm which focus (large/small) faults in IF Remote Adjust, then measure filament resistance at the tube terminals — an open fine-focus filament (OL) condemns the X-ray tube.
Toshiba / Canon Medical CT systems that use the XC (X-ray Controller) exposure subsystem, including the Alexion Access (TSX-033A), Activion16 (TSX-031A) and the Aquilion family. Field note: a persistent Error in XC that appears only on the small (fine) focus commonly points to an open fine-focus filament in the X-ray tube — in IF Remote Adjust the fault tracks the focus selection (clean on large focus, faulting on small), and a resistance check at the tube terminals shows the fine filament open (OL) while the coarse filament reads a fraction of an ohm.
Full troubleshooting for B2640000
The complete diagnosis, cause tree, and step-by-step remedy for the X-ray controller (XC) / exposure subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated X-ray controller (XC) / exposure codes