Information / dialogReconstruction (RECON) unit

00000002RECON unit timed out — restart the reconstruction queue

Canon Medical (Toshiba) CT error code, reported by the Reconstruction (RECON) unit.

What 00000002 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 00000002 indicates: RECON unit timed out — restart the reconstruction queue. The code originates in the Reconstruction (RECON) unit section of the Reconstruction (RECON) unit. Severity level 1 (Information / dialog).

Reported action: Manually restart the Reconstruction queue, exactly as the OEM system-software message instructs. If the RECON unit keeps timing out after the restart, check its own status and connection before resuming scanning. If what actually showed was only "Localizer is lighting now.", that is the unrelated benign status message and needs no service action..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. This code number is shared across the 1:SS Display Message, 20:Device PIP and 80:Device APPL tables in errmsg.eng, whose OEM texts are "Localizer is lighting now.", "Error in RECON unit (Time out)." and "Error in system software. Manualy restart Reconstruction queue." — the RECON time-out is the real fault, and the system-software text explicitly instructs restarting the reconstruction queue manually; "Localizer is lighting now." is the routine status use of the same code number.

Full troubleshooting for 00000002

The complete diagnosis, cause tree, and step-by-step remedy for the Reconstruction (RECON) unit subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Reconstruction (RECON) unit codes

Related service procedures

Discuss 00000002 with other field engineers in the FSELIB forum