Replace disk array units (BSX74-2035E) — disk drive, power supply, and fan units
Remove and install the field-replaceable sub-units of the BSX74-2035E raw data disk array (RDD) in the console REC BOX: the disk drive unit, the power supply unit, and either fan unit (FAN1/FAN2) — each accessed and serviced independently via its own carrier or screws. This task covers the mechanical R&R steps only; for the array's outline, specifications, internal daisy-chain wiring, LED status t
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Disk drive unit (Fibre-channel HDD, in carrier) | n/a in source | 1 (or more, per defective unit(s)) — fits slots #1-#5 only |
| Power supply unit (in carrier) | n/a in source | 1 |
| Fan unit (FAN1 and FAN2, independently removable) | n/a in source | 1 or 2 (each fan unit serviced independently) |
| Securing screws, fan unit | n/a in source | 3 per fan unit |
Safety
Hazard: Subjecting the disk drive unit to shock while it is removed or inserted
Consequence: Equipment damage — the disk drive is particularly sensitive to shock; source gives no personal-injury language for this precaution
Avoidance: Handle the disk drive unit (and power supply unit) without shock throughout removal and insertion
Hazard: Changing the slot arrangement of the disks (e.g., swapping order during service) without recording which slot a disk came from, then reusing that disk
Consequence: The stored data on the disk array becomes invalid if the arrangement of the disks changes
Avoidance: Before removing a disk you intend to reuse, write down the number of the slot it was set in, so it can be remounted in the correct slot
Full step-by-step for Replace disk array units (BSX74-2035E) — disk drive, power supply, and fan units
All 10 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Console guides
Source: 2D201-113EN*J §4.6.9.5 'Procedures for removing/installing the disk drive unit/power supply unit' and §4.6.9.6 'Procedures for removing/installing the fan unit' (pp.468-471) — BSX74-2035E raw data disk array (RDD) variant. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.