00000005 — RECON unit error — incorrect parameter set / Stack Scan aborted
Canon Medical (Toshiba) CT error code, reported by the Reconstruction (RECON) unit.
What 00000005 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 00000005 indicates: RECON unit error — incorrect parameter set / Stack Scan aborted. The code originates in the Reconstruction section of the Reconstruction (RECON) unit. Severity level 0 (Information / dialog).
Reported action: Note which of the two messages was displayed. For the RECON unit error, review the reconstruction parameter set of the job that failed and reconstruct again from the raw data. For the aborted Stack Scan, repeat the scan and check the system log for a preceding fault; call service if the code returns..
Listed for Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003, under the 20:Device PIP and 80:Device APPL tables. Two OEM messages share this code value: "Error in RECON unit ( incorrect paramter set)." (OEM spelling) and "Aborted Stack Scan."
Full troubleshooting for 00000005
The complete diagnosis, cause tree, and step-by-step remedy for the Reconstruction subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Reconstruction codes
Related service procedures
Source: errmsg.eng — 20:Device PIP / 80:Device APPL tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.