Pular para o conteúdo principal
Information / dialogReconstruction (RECON) unit

00000001RECON unit ended abnormally, reconstruction queue full

Canon Medical (Toshiba) CT error code, reported by the Reconstruction (RECON) unit.

What 00000001 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 00000001 indicates: RECON unit ended abnormally, reconstruction queue full. The code originates in the Reconstruction (RECON) unit section of the Reconstruction (RECON) unit. Severity level 1 (Information / dialog).

Reported action: Check the RECON (reconstruction) unit's own status and the system log for the point where processing stopped, and if the queue is also reported full, work through and clear pending reconstruction jobs so there is room before submitting new ones. If what actually showed on the console was only "Table is free now.", that is the unrelated benign status message and needs no service action..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. This code number is shared across the 1:SS Display Message, 20:Device PIP and 80:Device APPL tables in errmsg.eng, whose OEM texts are "Table is free now.", "Error in RECON unit (abnormal end)." and "Reconstruction queue FULL." — the RECON-unit and queue-full messages are what is worth troubleshooting; "Table is free now." is the routine status use of the same code number.

Full troubleshooting for 00000001

The complete diagnosis, cause tree, and step-by-step remedy for the Reconstruction (RECON) unit subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Reconstruction (RECON) unit codes

Related service procedures

Discuss 00000001 with other field engineers in the FSELIB forum

Source: errmsg.eng — 1:SS Display Message / 20:Device PIP / 80:Device APPL tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.