ProcedureConsolePreventive MaintenanceClean

Clean the console mouse

Clean the console's PS/2 wheel mouse when its movement becomes abnormal: open the ball holder, wipe the ball and its 3 internal rollers, reassemble, and verify. This is one of the three Operation-section components alongside the monitor and the keyboard and control PWB (see Related) — use this page specifically for mouse cleaning/replacement; for the monitor or keyboard, go to their own reference

Applies to
Activion16 (TSX-031A). No model- or revision-specific restriction is stated in source for the console mouse (PS/2 standard mouse with wheel, mini-DIN 6-pin connector).
Personnel
1 technician · "Field service engineer (Console — Operation section: monitor/keyboard/mouse); no special tools or disassembly beyond the mouse itself"
Steps
4 steps
Source
2D201-113EN*J §4.5.4 'Mouse' (pp.424-425)

Tools & parts

ItemPart numberQty
Soft cloth, moistened with neutral detergent and wrung outn/a in sourceas needed
Replacement mouse (PS/2 standard mouse with wheel, mini-DIN 6-pin connector)n/a in source1 (only if the mouse is still abnormal after cleaning, or is damaged)

Safety

NOTICE

Hazard: Using liquids such as detergents to clean the ball and the 3 rollers inside the mouse

Consequence: Not stated explicitly in source (equipment-quality concern only — FSELIB assigns NOTICE, not CAUTION, per the platform's ANSI Z535 property-damage tier)

Avoidance: Do not use liquids such as detergents on the ball/rollers. Remove only dirt and dust, and take care to prevent dust from remaining inside the mouse.

NOTICE

Hazard: Wiping the mouse's outer surface with alcohol or organic solvents

Consequence: Not stated explicitly in source (equipment-quality concern only — FSELIB assigns NOTICE, not CAUTION)

Avoidance: Wipe the outer surface only with a soft cloth that is moistened with neutral detergent and wrung out. Do not use alcohol or organic solvents.

NOTICE

Hazard: Connecting a mouse other than the one provided in the standard configuration, or one approved by Nasu Operations

Consequence: Not stated in source (no injury or equipment-damage outcome is given; treat as an approved-hardware compatibility requirement)

Avoidance: The mouse must be the one provided in the standard configuration, or one that has been approved by Nasu Operations. Do not connect any other mouse to the system.

Full step-by-step for Clean the console mouse

All 4 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Console guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §4.5.4 'Mouse' (pp.424-425). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.