00000003 — RECON unit failed to start — delete the reconstruction queue
Canon Medical (Toshiba) CT error code, reported by the Reconstruction (RECON) unit.
What 00000003 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 00000003 indicates: RECON unit failed to start — delete the reconstruction queue. The code originates in the Reconstruction (RECON) unit section of the Reconstruction (RECON) unit. Severity level 1 (Information / dialog).
Reported action: Manually delete the Reconstruction queue entry, exactly as the OEM system-software message instructs, then retry. If the RECON unit still fails to start, check its own status and connection. If what actually showed was only "Couch is in local mode (GTS Local).", that is the unrelated benign status message and needs no service action..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. This code number is shared across the 1:SS Display Message, 20:Device PIP and 80:Device APPL tables in errmsg.eng, whose OEM texts are "Couch is in local mode (GTS Local).", "Error in RECON unit (failed to start)." and "Error in system software. Manualy delete Reconstruction queue." — the RECON unit failing to start is the real fault, and the system-software text explicitly instructs deleting the reconstruction queue manually; "Couch is in local mode (GTS Local)." is the routine status use of the same code number.
Full troubleshooting for 00000003
The complete diagnosis, cause tree, and step-by-step remedy for the Reconstruction (RECON) unit subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Reconstruction (RECON) unit codes
Related service procedures
Source: errmsg.eng — 1:SS Display Message / 20:Device PIP / 80:Device APPL tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.