Replace the host PC unit
Remove and replace the host PC unit — the computer mounted in the CPU BOX cabinet that controls the entire Activion16 system and performs image-processing operations such as 3D processing — on either of its two documented hardware generations, PX74-08766 Gr.4 or PX74-09549 Gr.3. The removal procedure is identical for both generations.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Host PC unit assembly, PX74-08766 Gr.4 | PX74-08766 | 1 |
| Host PC unit assembly, PX74-09549 Gr.3 | PX74-09549 | 1 |
| USB LAN module (option hardware/license key) | n/a in source | 1 |
Safety
Hazard: Host PC unit not held securely while being pulled out toward the front (Step 4)
Consequence: The unit can fall; its front may hit the floor — an equipment-damage concern, with no personal-injury language in source
Avoidance: Hold the unit securely throughout the pull-out motion
Hazard: Replacing (discarding) the USB LAN module when replacing the host PC unit itself
Consequence: Loss of the option hardware/license key that identifies the registered user on this system
Avoidance: Do not replace the USB LAN module — transfer the existing module from the old host PC unit to the replacement unit
Full step-by-step for Replace the host PC unit
All 6 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Console guides
Source: 2D201-113EN*J §4.6.3.1 'Host PC unit' (p.431 removal procedure; p.432 left-side-cover access; p.433 USB LAN module notice) — PX74-08766 Gr.4 hardware generation; identical removal-procedure text repeated at §4.6.4.1 (p.438) for the PX74-09549 Gr.3 generation. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.