Pular para o conteúdo principal
ProcedureConsoleReplacement/FRUReplace

Replace the host PC unit

Remove and replace the host PC unit — the computer mounted in the CPU BOX cabinet that controls the entire Activion16 system and performs image-processing operations such as 3D processing — on either of its two documented hardware generations, PX74-08766 Gr.4 or PX74-09549 Gr.3. The removal procedure is identical for both generations.

Applies to
Activion16 (TSX-031A). Covers both Host PC unit hardware generations documented in this manual: PX74-08766 Gr.4 (§4.6.3, pp.431-437) and PX74-09549 Gr.3 (§4.6.4, pp.438-442). The removal-procedure text is identical between the two generations — only internal composition/specifications differ (see the per-generation reference atoms, Related).
Personnel
1 technician · Field service engineer (CT console / Navibox hardware)
Steps
6 steps
Source
2D201-113EN*J §4.6.3.1 'Host PC unit' (p.431 removal procedure; p.432 left-side-cover access; p.433 USB LAN module notice) — PX74-08766 Gr.4 hardware generation; identical removal-procedure text repeated at §4.6.4.1 (p.438) for the PX74-09549 Gr.3 generation

Tools & parts

ItemPart numberQty
Host PC unit assembly, PX74-08766 Gr.4PX74-087661
Host PC unit assembly, PX74-09549 Gr.3PX74-095491
USB LAN module (option hardware/license key)n/a in source1

Safety

NOTICE

Hazard: Host PC unit not held securely while being pulled out toward the front (Step 4)

Consequence: The unit can fall; its front may hit the floor — an equipment-damage concern, with no personal-injury language in source

Avoidance: Hold the unit securely throughout the pull-out motion

NOTICE

Hazard: Replacing (discarding) the USB LAN module when replacing the host PC unit itself

Consequence: Loss of the option hardware/license key that identifies the registered user on this system

Avoidance: Do not replace the USB LAN module — transfer the existing module from the old host PC unit to the replacement unit

Full step-by-step for Replace the host PC unit

All 6 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Console guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §4.6.3.1 'Host PC unit' (p.431 removal procedure; p.432 left-side-cover access; p.433 USB LAN module notice) — PX74-08766 Gr.4 hardware generation; identical removal-procedure text repeated at §4.6.4.1 (p.438) for the PX74-09549 Gr.3 generation. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.