Replace disk drive, power supply, or fan units in the BSX74-1900E raw data disk array
Remove and install the field-replaceable disk drive units, the power supply unit, and the fan units of the BSX74-1900E raw data disk array (RDD) — the JBOD dual-loop fiber-channel array that stores raw scan data, reconstruction functions, and correction data sets for the console.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Disk drive unit (Fibre-channel HDD, mounted in its carrier) | n/a in source | up to 5 (slots #1 to #5) |
| Power supply unit (mounted in its carrier) | n/a in source | 1 |
| Fan unit | n/a in source | 2 |
Safety
Hazard: Subjecting the disk drive unit or power supply unit to shock while it is being removed or inserted (Precautions, p.463) — source specifically calls out the disk drive as particularly shock-sensitive
Consequence: Equipment damage to the shock-sensitive disk drive unit; source gives no personal-injury language for this step
Avoidance: Do not subject the unit to shocks during removal or insertion
Hazard: Reinstalling a removed disk drive unit in a different numbered slot than the one it was removed from
Consequence: The stored data becomes invalid if the arrangement of the disks changes
Avoidance: Before removing a disk drive unit that will be reused, write down the number of the slot it came from, so it can be remounted in the same slot
Full step-by-step for Replace disk drive, power supply, or fan units in the BSX74-1900E raw data disk array
All 9 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Console guides
Source: 2D201-113EN*J §4.6.8.5 'Procedures for removing/installing the disk drive unit/power supply unit' (pp.461-463) and §4.6.8.6 'Procedures for removing/installing the fan unit' (p.464) — BSX74-1900E raw data disk array. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.