Pular para o conteúdo principal
Canon Medical·🔬CT Scanner

Activion16 (TSX-031A) Service Manual

FSELIB authored service atoms

FSELIB authored service atoms for the Canon Medical (Toshiba) Activion16 CT scanner: gantry, couch, X-ray generator, DAS/detector, data transmission and console. Web-first, real figures at 300 dpi with PDF-fallback for oversized diagrams. This service manual covers installation, calibration, troubleshooting and maintenance procedures for the Canon Medical Activion16 (TSX-031A).

Also known as: Toshiba, Canon, Activion, Activion16, TSX-031A

Start free trial →

Access 534+ technical service manuals — 7-day free trial, no credit card required.

534+ manuals · 7+ manufacturers · CT, MRI, Mammography, X-Ray & more

Technical Specifications

ManufacturerCanon Medical
Equipment TypeCT Scanner
Part Number
2D201-112/113EN
RevisionFSELIB 2026-07
Models Covered
Activion16TSX-031A

Table of Contents

From actual manual
1Adjust the gantry positioning projectors (side, slant, and median)
Start free trial →
2Adjust the gantry rotation belt tension
Start free trial →
3Adjust the gantry rotation speed (servo amplifier, VR1)
Start free trial →
4Adjust the gantry servo-amplifier offset
Start free trial →
5Adjust the gantry tilt-stop angles (encoder calibration)
Start free trial →
6Check the gantry rotation speed
Start free trial →
7Check the gantry rotation speed voltage from the KGTSM PWB to the servo amplifier
Start free trial →
8Check the safety circuit on the KGTSM PWB
Start free trial →
9Clean the gantry exterior covers, Mylar ring, and mat switch
Start free trial →
10Control flow — from operation request to gantry/couch motion and acquisition data (§3.3.5)
Start free trial →
11Control section composition — the seven units behind gantry/couch control
Start free trial →
12Control section outline — where it lives and what triggers it
Start free trial →
13Control section — system block diagram and where each block sits
Start free trial →
14Diagnose gantry tilt faults — tilt disabled or noisy during tilt operation
Start free trial →
15GTS/KGTSM gantry-rotation error (b1360000 family) — decode & isolate
Start free trial →
16Gantry & couch — additional data words reference
Start free trial →
17Gantry configuration — the four assemblies and how they fit together
Start free trial →
18Gantry mechanical troubleshooting — tilt, rotation, slip ring, wedge, slit (by symptom)
Start free trial →
19Gantry rotation interlock — what the cover switch and the Enable switch actually do
Start free trial →
20Gantry service — general safety precautions (read this first)
Start free trial →
21Gantry — fastening torques reference (important mounting assemblies)
Start free trial →
22Gantry — servo motor offsets & rotation-speed reference
Start free trial →
23Gantry/couch control section — outline: the PWBs involved and how you diagnose them
Start free trial →
24Interlock — how the Activion16 stops itself, and in what order
Start free trial →
25KGTSM control PWB — VR and test-point (TP) reference
Start free trial →
26KGTSM control PWB — connectors, LEDs & test points reference
Start free trial →
27KGTSM switch bank — normal-use settings reference
Start free trial →
28KGTSM terminal messages — collection, timing & full message catalog
Start free trial →
29KGTSM timer self-check — timer error (m10 bit D2) — read & isolate
Start free trial →
30KGTSM — DSW4 and DSW11 system-setting switches, explained
Start free trial →
31KGTSP control PWB — JP1 lamp-mode jumper & VR1/VR2 mic-gain trim
Start free trial →
32Main frame unit — theory of operation (rotation, power & signal transfer)
Start free trial →
33Activion16 system architecture — the whole machine at a glance
Start free trial →
34Manually tilt the gantry after a tilt-drive malfunction (emergency)
Start free trial →
35Open or close the gantry front cover
Start free trial →
36Operate the KGTSM maintenance panel
Start free trial →
37Optical system operation — how the wedge and slit motors find, hold, and time out on position
Start free trial →
38Perform the checks after replacing the KGTSM PWB, servo amplifier, or servo motor
Start free trial →
39Release the gantry rotation brake to turn the rotation section by hand (emergency)
Start free trial →
40Remove and mount the gantry rear-upper covers (left and right)
Start free trial →
41Remove and refit the Activion16 gantry top cover
Start free trial →
42Remove and refit the gantry Mylar ring
Start free trial →
43Remove and refit the gantry dome cover
Start free trial →
44Remove or mount the gantry side cover (right or left)
Start free trial →
45Replace a gantry laser-positioning projector lamp
Start free trial →
46Replace the gantry optical system (slit/wedge assembly)
Start free trial →
47Replace the gantry rotation motor
Start free trial →
48Replace the gantry tilt gas spring
Start free trial →
49Service the gantry tilt drive: lubricate the tilt detection unit and replace the power cylinder
Start free trial →
50Set up and verify the gantry rotation servo amplifier
Start free trial →
51Slit section won't move, won't return, or is noisy — troubleshoot and repair
Start free trial →
52Stand unit — how the gantry tilt drive works
Start free trial →
53Targets of control — what the gantry/couch control section actually governs
Start free trial →
54Troubleshoot da seção do anel coletor (slip ring) — escova gasta, ruído e descarga (Gantry)
Start free trial →
55Troubleshoot the Activion16 gantry rotation section
Start free trial →
56Use the KGTSM maintenance mode from a PC terminal
Start free trial →
57Wedge section troubleshooting — filter won't operate, abnormal noise, or unstable wedge block
Start free trial →
58Adjust the couch descent (lowering) speed — solenoid throttle valves L21/L24
Start free trial →
59Adjust the couch height display
Start free trial →
60Adjust the couch-top longitudinal belt tension
Start free trial →
61Adjust the couch-top supporting rollers (height, level, and sway)
Start free trial →
62Change the patient couch vertical stop positions
Start free trial →
63Check the couch-top (horizontal) travel speed
Start free trial →
64Couch lower covers — the flexible cover and base covers
Start free trial →
65Couch main frame unit — theory of operation (the couch top's structural chassis)
Start free trial →
66Couch slide servo amplifier — control theory, LED/alarm decode, PANATERM tool & parameters
Start free trial →
67Couch top — construction and how it's secured
Start free trial →
68Couch upper covers — disassembly, naming, and mounting/removal procedures
Start free trial →
69Couch — routing of the internal cables at the linkage unit (work procedure)
Start free trial →
70Couch-top moving frame — how the couch top slides and levels
Start free trial →
71Enable or disable couch-top sliding fine adjustment (KGTSM `adj`)
Start free trial →
72Horizontal (longitudinal) couch motion speed — servo amplifier note
Start free trial →
73Jack up the patient couch when it cannot move up from the DOWN limit (emergency)
Start free trial →
74Longitudinal movement drive unit — how the couch top moves IN and OUT
Start free trial →
75Manually lower the patient couch after a power/fuse failure (emergency)
Start free trial →
76PM: Inspect and maintain the couch clutch/brake assembly (dry type)
Start free trial →
77Partially disassemble the patient couch for narrow carrying-in routes (CBTB-018A only)
Start free trial →
78Patient couch control troubleshooting — KGTSM / KGTSP / slide servo amplifier faults
Start free trial →
79Patient couch mechanical section — scope and the CBTB-018A / CBTB-018B split
Start free trial →
80Patient couch structure and operation — the SLIDE/HEIGHT/SI-to-KGTSM map
Start free trial →
81Patient couch troubleshooting — vertical lift, horizontal slide, rollers, and noise (by symptom)
Start free trial →
82Replace the couch horizontal (longitudinal) drive unit
Start free trial →
83Replace the patient couch hydraulic unit
Start free trial →
84Replace the patient couch's flexible cover
Start free trial →
85Set the couch autoslide movement range (KGTSM console)
Start free trial →
86Set the couch horizontal movement absolute-value control data (KGTSM console)
Start free trial →
87Set the couch vertical movement upward stop position (Auto-set mode, KGTSM console)
Start free trial →
88Set the couch-sliding zero-clear switch hold time (KGTSM `ctm`)
Start free trial →
89Vertical movement drive section — how the Activion16 couch moves up and down
Start free trial →
90Vertical movement mechanism — the linkage that lifts the couch
Start free trial →
91Adjust the X-ray tube alignment (front/rear and left/right)
Start free trial →
92Gantry X-ray tube position information — encoder counts and where they reset
Start free trial →
93Replace the 4-MHU X-ray tube starter
Start free trial →
94Replace the 42-kW HFG (CXXG-010A)
Start free trial →
95Replace the X-ray tube
Start free trial →
96X-ray beam control: how the wedge filter and slit unit shape the beam
Start free trial →
97AC/CONTROL section: how it powers and locally monitors the inverter circuit
Start free trial →
98Adjust the If (filament) value remotely from the console
Start free trial →
99Anode heat capacity: how OLP calculates it, enforces the limit, and survives a power cycle
Start free trial →
100Check the X-ray generator power-on and power-off sequence
Start free trial →
101Check the X-ray tube current with the tube ammeter (MA-1201D)
Start free trial →
102Check the X-ray tube voltage with the tube voltage meter (kV-201 bleeder)
Start free trial →
103Handle the high-voltage cable — removal, cleaning, and silicone-oil reassembly (CLAYMOUNT)
Start free trial →
104INV/HV section operating principle — how the inverter and HV tank turn 300 VDC into tube voltage
Start free trial →
105OLP management functions — what the X-ray tube system protects and how it decides
Start free trial →
106Replace the X-ray generator fuse
Start free trial →
107Retrieve and decode the XC PWB error log (direct RS-232C read + system-software XL30 console log)
Start free trial →
108Run a single-unit X-ray exposure from the console
Start free trial →
109Short-term monitoring of the nominal rating — OLP protection for repeated same-kV/mA exposures
Start free trial →
110Short-term monitoring of the rating — protecting the focus during continuous radiation
Start free trial →
111Starter ST-7011,B (CXB-400C anode drive) — composition, ratings, signals & interlock reference
Start free trial →
112Starter ST-7013 (CXB-400C anode drive) — composition, ratings, signals & interlock reference
Start free trial →
113Troubleshoot starter abnormalities (ST-7011,B / ST-7013) — E1-E8, inverter display codes, shaft lock release
Start free trial →
114Troubleshoot the Activion16 X-ray generator main circuit (9 fault flowcharts)
Start free trial →
115X-ray Generator/HV — service precautions (floating section, power cable, rotation, capacitor discharge)
Start free trial →
116X-ray HV generator: E00-E90 error codes, PWB setup, and the 42-kW parts list
Start free trial →
117X-ray high-voltage generator — composition (AC/CONTROL + INV/HV + interconnecting cables)
Start free trial →
118X-ray high-voltage generator — operating principle
Start free trial →
119Check the X-ray tube aging (warm-up and seasoning) schedule
Start free trial →
120Check the X-ray tube getter aging (heat exchanger pushbutton)
Start free trial →
121Clean the heat exchanger fan (annual PM)
Start free trial →
122Connect the replacement X-ray tube and run the operation check
Start free trial →
123X-ray tube system configuration — tube unit and heat exchanger as one sealed assembly
Start free trial →
124X-ray tube troubleshooting — discharge, oil leakage, tube interlock, noise, and no tube current (by symptom)
Start free trial →
125X-ray tube — specifications reference (CXB-400C ratings + cooling curve)
Start free trial →
126X-ray system block diagram — how the generator, control, and tube blocks fit together
Start free trial →
127X-ray system configuration — the HFG, the tube, and the starter
Start free trial →
128X-ray system — specifications reference (voltage, current, time, focus, line capacity)
Start free trial →
129X-ray system — tools and instruments reference
Start free trial →
130AJ DAS I/O cable connector pinout — CNN1, CNN2 (High speed), CNN3 (Low speed)
Start free trial →
131AJ DAS test data patterns — IntegA / IntegB reference tables
Start free trial →
132AJ-CONTROL PWB — DIP switches, LEDs, and DAS-fan monitor (§2.14.2a)
Start free trial →
133AJ-CONTROL PWB — OPCONTA interface signals and the DAS acquisition sequence (§2.14.2b)
Start free trial →
134AJ-CONVERTER PWB — function and channel/integrator architecture (DAS)
Start free trial →
135Check the DAS supply voltage (AJ-CONTROL / AJ-CONVERTER PWBs)
Start free trial →
136Check the DAS supply voltage (Control-16 / Converter-16 PWBs)
Start free trial →
137Control-16 PWB — OPCONTA interface signals and the DAS acquisition sequence (§2.6.2b)
Start free trial →
138Control-16 PWB — switches, LEDs, and connectors (DAS)
Start free trial →
139Converter-16 PWB — function and channel/integrator architecture (DAS)
Start free trial →
140DAS I/O cable connector pinout — CNN1, CNN2 (High speed), CNN3 (Low speed)
Start free trial →
141DAS gain-dependent abnormality — reading the GAIN/odd-even-view pattern (§2.8.5)
Start free trial →
142DAS integrator-dependent abnormality — odd/even-view fault pattern (Integ-A / Integ-B)
Start free trial →
143DAS power supply — +5.7 V / +3.3 V / +24 V rails, terminals, and OV/OC protection reset
Start free trial →
144DAS test data patterns — IntegA / IntegB reference tables
Start free trial →
145DAS — before repair (isolate power, and replace rather than field-repair)
Start free trial →
146DAS — composition and channel-count math (AJ-CONTROL + AJ-CONVERTER PWBs)
Start free trial →
147DAS — composition and channel-count math (Control-16 + Converter-16 PWBs)
Start free trial →
148DAS — data acquisition section components (CDAS-035A/1A and CDAS-035A/2A "AJ")
Start free trial →
149DAS — data acquisition section configuration table (CDAS-035A/1A and CDAS-035A/2A "AJ")
Start free trial →
150DAS — gantry rear-panel layout drawing (component identification)
Start free trial →
151DAS — identify which CDAS hardware version is fitted (CDAS-035A/1A vs /2A)
Start free trial →
152DAS — notices on work (service precautions before touching the data acquisition system)
Start free trial →
153DAS — tools and instruments reference
Start free trial →
154DAS/detector troubleshooting — slice thickness-dependent (row-dependent) abnormality
Start free trial →
155Data acquisition faults — cause × symptom matrix and 4 isolation flowcharts (DAS)
Start free trial →
156Data flow — from the main/REF detector to the MHR2-FC PWB (§2.3)
Start free trial →
157Replace the DAS PWB (Converter-16 / Control-16)
Start free trial →
158Run the DAS DCA test procedures (X-ray output, offset data, test pattern checks)
Start free trial →
159Troubleshoot Activion16 DAS data-acquisition image faults — cause matrix (AJ-CONTROL / AJ-CONVERTER)
Start free trial →
160Detector-channel-to-DAS-channel correspondence — Converter-16 PWB slot mapping (Detector / DAS)
Start free trial →
161Main detector — 800-channel x 16-row solid-state array, bundled to 712 channels before the DAS (§2.4)
Start free trial →
162Main detector, REF detector, and DAS — connection diagram (§2.5.1)
Start free trial →
163REF detector — solid-state reference sensor for view-direction X-ray dose variation
Start free trial →
164Replace the REF detector
Start free trial →
165Replace the main detector
Start free trial →
166Set up the temperature controller after replacement (main detector)
Start free trial →
167Check gantry rotation and MUDAT data transfer after working on the MUDAT
Start free trial →
168Check the MUDAT transmission-LED luminescence (ROT and SOT sides)
Start free trial →
169DTU — MUDAT physical layout (rotation and stationary rings, rear view)
Start free trial →
170Data transfer section error-identification test tools (DAS / Console / ROTSOT / SOTROT / OPCONTA / GCIFA)
Start free trial →
171Data transfer section errors — the 25 SOTROT/ROTSOT codes (reference)
Start free trial →
172Data transmission unit — what it does and the 19-term glossary behind the rest of this chapter
Start free trial →
173Disassemble the MUDAT (rotation and stationary section PWBs)
Start free trial →
174Functions of the MUDAT — how data crosses the rotating joint
Start free trial →
175GCIFA PWB — DIP switches, fuse, and board layout
Start free trial →
176GCIFA PWB — LED list (status, error, serial-comm, comm-signal)
Start free trial →
177GCIFA console test patterns — 16-segment expected-value reference
Start free trial →
178GCIFA-D2 PWB troubleshooting flowchart (Data Transmission)
Start free trial →
179GCIFA-D2 PWB — DIP switch, connectors, LEDs, and startup status
Start free trial →
180IF unit operating principle: OPCONTA, GCIFA, GCIFA-D2, and channel switching (§3.3.2, items 1-4)
Start free trial →
181IF unit — external appearance (where OPCONTA and GCIFA physically sit)
Start free trial →
182IF unit — what OPCONTA and GCIFA each do in the data-transfer path
Start free trial →
183MUDAT optical data-transfer section - sub-unit composition & PWB reference
Start free trial →
184Malfunction of DAS data communication — cause investigation (streak/ring artifacts)
Start free trial →
185Malfunction of control system data communication — MUDAT ROTSOT/SOTROT fault tree (Data Transmission)
Start free trial →
186OPCONTA PWB — DIP switches, fuses, and board layout
Start free trial →
187OPCONTA PWB — LED list (optical-control + MUDAT/XC/FIFO/TX signals)
Start free trial →
188Output GCIFA console test data and read the transmission-error display panel
Start free trial →
189Set the GCIFA tube-position switches for a Disable (non-rotation) scan
Start free trial →
190System diagram and operating principle — the MUDAT's light-emitting and light-receiving circuitry
Start free trial →
191Adjust the SS/ADI4 audio (gantry/console speaker and microphone VRs)
Start free trial →
192Boot-up and scanning control/data flow (§4.6.2)
Start free trial →
193CPU BOX power CONT — fuses, PS-rail standards & DIP SW1 gantry-reset-time setting (PX74-07907 / PX74-09756)
Start free trial →
194Clean the console mouse
Start free trial →
195Console — 1000BASE HUB reference
Start free trial →
196Console — CNN PANEL (CPU BOX) connector signal pinouts
Start free trial →
197Console — CPU BOX POWER CONT internal wiring: PX74-07907 vs PX74-09756 (reference)
Start free trial →
198Console — CPU BOX internal layout
Start free trial →
199Console — CPU BOX power system connection diagram (reference)
Start free trial →
200Console — CPU BOX power transformer (reference)
Start free trial →
201Console — Host PC unit reference (Gr.3, PX74-09549)
Start free trial →
202Console — Host PC unit reference (Gr.4, PX74-08766)
Start free trial →
203Console — Keyboard and control PWB (reference)
Start free trial →
204Console — Navibox external connection cable map (CNN panels)
Start free trial →
205Console — REC BOX Power CONT (board photo, component locations)
Start free trial →
206Console — REC BOX power system connection diagram (reference)
Start free trial →
207Console — Raw data disk array (BSX74-2035E): outline, internal configuration, LEDs, connection & setup
Start free trial →
208Console — SS/ADI4 PWB reference (switches, LEDs, connectors)
Start free trial →
209Console — configuration and external view (reference)
Start free trial →
210Console — list of fuses (reference)
Start free trial →
211Console — operation-section monitor (reference)
Start free trial →
212Console — signal system connection diagram (reference)
Start free trial →
213MHR PWB — outline, specifications, LEDs and settings (reference)
Start free trial →
214MHR data processing architecture and flow — internal nodes, preprocessing path, and reconstruction chain
Start free trial →
215Navibox configuration and console block diagram
Start free trial →
216Power ON/OFF sequence — how the Navibox power controller stages console and gantry power
Start free trial →
217REC BOX connector signal pinouts — DAS, CONT FC-AL, SCONT1, SCONT2
Start free trial →
218REC BOX — internal component layout reference
Start free trial →
219RTM-D PWB (PX74-08763) — function and LED list
Start free trial →
220RTM-M PWB (BSX74-1935 / BSX74-2245) — connectors, switches, LEDs & shorting-pin reference
Start free trial →
221Raw data disk array (BSX74-1900E) — outline, architecture, LEDs & connection reference
Start free trial →
222Raw data disk array — JBOD by Avid (Medea): configuration, LEDs, wiring & disk removal
Start free trial →
223Replace disk array units (BSX74-2035E) — disk drive, power supply, and fan units
Start free trial →
224Replace disk drive, power supply, or fan units in the BSX74-1900E raw data disk array
Start free trial →
225Replace the host PC unit
Start free trial →
226Run the hwtest malfunction diagnosis program (Console)
Start free trial →
227Run the independent MHR PWB test from an RTM terminal (diagnostics)
Start free trial →
228Clean the large-current slip ring
Start free trial →
229Gantry power-supply sequence — how MC111/MC112/MC113 turn the gantry on and off
Start free trial →
230Gantry — list of fuses (reference)
Start free trial →
231Replace the large-current slip ring brushes
Start free trial →
232Reset the patient couch's overcurrent relays (horizontal and vertical movement motors)
Start free trial →
233Set the DAS timer (auto pre-heat of the data acquisition section)
Start free trial →
234Set the couch footswitch mode (KGTSM console)
Start free trial →
235ESD / static-electricity precautions for service work
Start free trial →

235 sections · Full content available after sign in

Error Codes — decoded

55 codes

Service Procedures & Guides

105 guides

View all 105 guides →

All Activion16 (TSX-031A) resources →Manuals, error codes & specs for this model

Frequently asked questions

Where can I access the Canon Medical Activion16 (TSX-031A) service manual?

The full Activion16 (TSX-031A) service manual is on FSELIB — the field-service engineering library for medical imaging. Sign in or start a 7-day free trial to open the complete manual: installation, calibration, troubleshooting, error codes and replacement procedures.

What does the Activion16 (TSX-031A) service manual cover?

It covers Adjust the gantry positioning projectors (side, slant, and median), Adjust the gantry rotation belt tension, Adjust the gantry rotation speed (servo amplifier, VR1), Adjust the gantry servo-amplifier offset, Adjust the gantry tilt-stop angles (encoder calibration), Check the gantry rotation speed and more for the Canon Medical Activion16 (TSX-031A) (CT Scanner).

Does the Activion16 (TSX-031A) manual include error codes?

Yes — 55 decoded error codes for the Activion16 (TSX-031A), each with its meaning, severity and the subsystem that reports it.

Is the Activion16 (TSX-031A) service manual free to download?

Access is by subscription, with a 7-day free trial. FSELIB hosts authored, web-first service documentation (searchable procedures, diagrams and error-code references) — not a scraped PDF.

Related Manuals

💬 Forum Discussions about this Model