Remove and mount the Mylar sheet (ring)

The Mylar sheet is the flexible ring that seals the gap between the gantry's dome cover and front cover — remove it to get at whatever lies behind that joint, and mount it back precisely (overlap angle, orientation, and contact with the dome all matter) so the joint stays sealed against fluid ingress.

Applies to
Aquilion 16 (TSX-101A) — Gantry mechanism, source volume 2D201-061
Personnel
1 technician · Basic — hand tools only (M4 screws, a scale, and optionally a bladed screwdriver); no electrical work or power-down is called out in source for this item (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
5 steps
Source
2D201-061E*H p.52-54

Tools & parts

ItemPart numberQty
M4 screws (2)n/a in source
Scale (as shown in Figure 6.1-4)n/a in source
Bladed screwdriver (alternative to the scale)n/a in source

Safety

NOTICE

Hazard: Mylar sheet not mounted securely / not in close contact with the dome cover, leaving a gap between the sheet and the cover

Consequence: Fluids such as blood or contrast medium can leak into the system through that gap, resulting in damage to the system or deterioration of image quality

Avoidance: Mount the Mylar ring securely and confirm close contact with the dome using the Verification step below before returning the gantry to service

Full step-by-step for Remove and mount the Mylar sheet (ring)

All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure