Open and close the gantry front cover
Open the gantry's front cover to reach the mechanism behind it — couch to the DOWN limit, four covers/rings off, the cover's own screws out, then swing the dome open and hook both support pipes in place — and close it again by running the same sequence in reverse.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Front under cover | n/a in source | |
| Mylar ring | n/a in source | |
| Side covers, left and right | n/a in source | |
| Front cover — bottom + side screws | n/a in source | |
| Supporting pipe, left | n/a in source | |
| Supporting pipe, right | n/a in source |
Safety
Hazard: Front cover, left in the open position, contacting the ceiling or the patient couch if the gantry is tilted while the cover is open
Consequence: The system may be damaged (source states an equipment-damage consequence only — no personal-injury language)
Avoidance: Be extremely careful not to allow the front cover to come into contact with the ceiling or the patient couch when the gantry is tilted with the front cover open; check clearance against Figure 6.3-2 before commanding a tilt while the cover is still open
Full step-by-step for Open and close the gantry front cover
All 10 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry — Covers & access guides
Source: 2D201-061E*H p.56-57. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.