Pular para o conteúdo principal
ProcedureGantry — Covers & accessClean

Clean the FRP covers and remove the wedge/slit optical system

Wipe down the gantry's FRP covers and Mylar ring, then — as a separate job that happens to share the same page in source — remove the wedge/slit optical system as a complete assembly with its base.

Applies to
Aquilion 16 (TSX-101A) — Gantry Mechanism volume, source 2D201-061, sections 8.7 (Cover Cleaning Method) and 8.8.1 (Wedge/slit section removal)
Personnel
1 technician · Basic for the cover cleaning (wipe with a damp cloth); Intermediate for the wedge/slit optics removal — isolating power at NFB2, engaging the rotation lock pin, and disconnecting four sensor/motor connectors near a motor that runs hot (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
7 steps
Source
2D201-061E*H p.115

Tools & parts

ItemPart numberQty
Clothn/a in source1
Mild detergent or warm watern/a in source
Rotation lock pinn/a in source1
Hand tools (screwdriver, wrench — size not specified)n/a in source1 set

Safety

CAUTION

Hazard: Touching the motor while removing the wedge/slit optical system

Consequence: The motor is hot and can cause burn injury

Avoidance: Do not touch the motor when working on the wedge/slit section

Full step-by-step for Clean the FRP covers and remove the wedge/slit optical system

All 7 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry — Covers & access guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-061E*H p.115. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.