Pular para o conteúdo principal
ProcedureGantry — Covers & accessRemove/Mount

Remove and mount the dome cover

Remove and mount the gantry's dome cover to reach the slip ring and brushes behind it, after killing all power to the slip ring as the DANGER on this page requires.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Basic — identifying and switching OFF breaker NFB2, then disconnecting one connector, one cable clamp, and four screws; no electrical work beyond operating the breaker, and no adjustment or calibration is called out in source
Steps
5 steps
Source
2D201-061E*H p.62

Tools & parts

ItemPart numberQty
Screwdriver (type not specified in source)n/a in source

Safety

DANGER

Hazard: High voltage present in the slip ring and the brushes

Consequence: Electric shock

Avoidance: Turn OFF all the breakers on the distribution board before removing the dome cover

Full step-by-step for Remove and mount the dome cover

All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry — Covers & access guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-061E*H p.62. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.