XC Maintenance Program: block test, exposure/running test, debug monitor and screens
Run the XC maintenance program's own block test, exposure test, running test, and cyclic debug-monitor read-out — the four built-in functional checks that confirm the SRU inverters, X-ray exposure path, and XC status reporting are all working, once the maintenance program is already connected and its test switch is set active.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| PC terminal running the XC maintenance program, already connected per subsections 8.4.1-8.4.3 | n/a in source |
Safety
Hazard: Reducing the KL (inverter drive) setting value below 200 while running the block test, in an attempt to raise a 0 kV or out-of-range reading
Consequence: The SRU inverter rating may be exceeded and the inverter may be damaged (equipment damage only — source labels this callout "CAUTION"; reclassified here as NOTICE per FSELIB convention because the stated consequence is equipment damage, not personal injury)
Avoidance: Do not reduce the KL setting value below 200 during the block test. If the output is still out of range at KL 200, do not keep lowering it — check the main inverter clock signal on the SRU filter board instead (see XC PWB reference, Related, which documents the identical KL>=200 floor from the PWB side).
Full step-by-step for XC Maintenance Program: block test, exposure/running test, debug monitor and screens
All 9 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.