XC Maintenance Program: Signal Show (kV/mA/If/HV waveforms)
Use the XC maintenance program's Signal Show screen to select, acquire and display live kV, mA, If and HV waveform traces — then set the display Range and Sampling so the trace reads clearly — the same measurement tool that the If-adjustment and VR3 kV-waveform-adjustment procedures both depend on.
Applies to
Aquilion 16 (TSX-101A) — X-ray & HV volume 2D201-064, section 8.4.11 "Signal Show" (source, p.110-112), part of the XC Maintenance Program (section 8.4). Requires the XC maintenance software V5.13 or later — source states plainly: "If the XC maintenance software version is earlier than V5.13, an error occurs when the Signal Show menu is used."
Personnel
Basic to Intermediate — select the correct kV/mA/If/HV check boxes before acquiring data, run Data Collect then Show in the right order, and set the Range (Start/End/Time) and Sampling (A/D conversion cycle time vs. display sampling frequency) so the trace being read is meaningful; also requires knowing the XC maintenance software must be V5.13 or later, since an older version errors out on this menu. No component-level fault isolation is asked for within this section itself — that judgment happens in the procedures (If adjustment, VR3 adjustment) that use this screen's output. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
1 step
Source
2D201-064E*H p.110-112
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| PC terminal running the XC maintenance software (V5.13 or later) | n/a in source |
Full step-by-step for XC Maintenance Program: Signal Show (kV/mA/If/HV waveforms)
All 1 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV (XC control) guides
If (filament current) adjustment procedure11 stepsXC Maintenance Program: outline, connection and menu1 stepsXC Maintenance Program: block test, exposure/running test, debug monitor and screens9 stepsSet up PWBs and adjust the kV waveform (VR3)4 stepsTroubleshoot XC operation, cable interlock and starter communication faults0 stepsReplace the XC PWB backup battery4 steps
View all Aquilion 16 (TSX-101A) guidesSource: 2D201-064E*H p.110-112. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.