X-ray tube/heat-exchanger leakage, rotor resistance and cooling-fan checks — 3.1-3.9
Work through nine periodic checks on the Aquilion 16's X-ray system — tube, heat exchanger, rubber hose, rotor, receptacle unit, and cooling fans — items **3.1** through **3.9** of the X-ray System periodic-maintenance table, with NFB120 turned OFF before you start any of them.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Heat exchanger | n/a in source | |
| Rubber hose | n/a in source | |
| 150 kV high-voltage cable | n/a in source |
Safety
Hazard: Leaving NFB120 (power to the rotation servo motor) ON while performing the checks in this X-ray System periodic-maintenance section
Consequence: The rotation section may rotate suddenly, resulting in injury or electric shock
Avoidance: Before starting operation on any item in this section, be sure to turn OFF NFB120. (Printed once, at the head of source subsection 5.3, X-ray System, ahead of item 3.1 — it governs the whole X-ray System table, of which items 3.1-3.9 are this atom's scope.)
Full step-by-step for X-ray tube/heat-exchanger leakage, rotor resistance and cooling-fan checks — 3.1-3.9
All 17 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV guides
Source: 2Y201-061E*E p.109-110. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.