X-ray HV: XC battery, charge/discharge, starter output, cables, SRU bolts, tube bellows — 3.14-3.19

Six periodic checks on the tail end of the Aquilion 16's X-ray system: back up memory and replace the XC PWB's lithium battery, verify the power-supply unit's charge/discharge timing, confirm starter output voltage, inspect XC/SRU/power-supply cables and connectors, torque-check the front SRU's molded connection bolts, and check the X-ray tube bellows for oil leakage — items **3.14** through **3.19** of the periodic X-ray-system inspection table.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Basic to Intermediate across items 3.14-3.19. Items 3.17 (cable/connector inspection) and 3.19 (bellows oil-leak check) are visual checks with no meter or tool. Item 3.14 is a basic component swap (one lithium battery) preceded by a memory backup — the full turn-off/swap/reset/readjust sequence lives in a separate, already-authored procedure (see Related). Item 3.15 (charge/discharge timing at TB2) and item 3.16 (starter output at TB5) are DC-voltage measurements against timed or fixed thresholds. Item 3.18 is a torque-wrench check restricted to the front-side SRU only. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
19 steps
Source
2Y201-061E*E p.113-115

Tools & parts

ItemPart numberQty
Lithium battery (ST Microelectronics M4T2B)**AVY005-01**
Multimetern/a in source
Torque wrenchn/a in source

Full step-by-step for X-ray HV: XC battery, charge/discharge, starter output, cables, SRU bolts, tube bellows — 3.14-3.19

All 19 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.