ProcedureX-ray & HVAdjust
X-ray tube alignment (front/rear and left/right)
Check, and if needed correct, the X-ray tube's front/rear (longitudinal) and left/right alignment — measuring each axis with the Alignment software against a fixed tolerance, then moving the tube by turning adjusting bolts on its mounting plate while reading a dial gauge — per Appendix 28.5 ("Tube Alignment") of the Installation manual, the method Section 18 of the main installation flow defers to.
Applies to
Aquilion 16 (TSX-101A) — Installation manual (source 2C201-160), Appendix 28.5 ("Tube Alignment"), p.188-192 — the method Section 18 ("X-ray Tube Alignment") of the main installation flow points to when it states only the pass/fail check.
Personnel
Intermediate — running the Alignment software's front/rear ("Scan to calculate") and left/right (0.75-s scan at the small focus, then the large focus) measurement functions, positioning a pin phantom so the center of the pin aligns with the gantry rotation center, and physically repositioning the X-ray tube: loosening the fixing bolts to free the tube mounting plate, then turning the adjusting bolts by a calculated number of turns while reading a dial gauge mounted on the oil cooler. Source states no OEM skill level for this appendix; this assessment is derived from the procedure's own content, not an OEM-stated skill level.
Steps
12 steps
Source
2C201-160E*F p.188-192
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Alignment software | n/a in source | 1 |
| Dial gauge | n/a in source | 1 |
| Pin phantom (with phantom holder, per the p.191 diagram) | n/a in source | 1 |
| Standard fastening tools | n/a in source | n/a in source |
Full step-by-step for X-ray tube alignment (front/rear and left/right)
All 12 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV guides
X-ray tube If adjustment (remote)11 stepsCheck phantom position, run the test scan and check tube alignment10 stepsX-ray HV: XC battery, charge/discharge, starter output, cables, SRU bolts, tube bellows — 3.14-3.1919 stepsX-ray tube/heat-exchanger leakage, rotor resistance and cooling-fan checks — 3.1-3.917 stepsGenerator troubleshooting table (phenomena and countermeasures)3 stepsRun the console malfunction-diagnosis program (hwtest)0 steps