Pular para o conteúdo principal
ProcedureX-ray & HVSetup & CalibrationAdjust

X-ray tube If adjustment (remote)

Remotely fine-tune the X-ray tube's filament current (If) through the XC maintenance program's If Adjustment screen — across the tube-voltage/focus-size/mA combinations selected — so tube current rises smoothly at exposure start, without connecting an oscilloscope to the XC PWB, and without the Over mA / Over If errors an incorrect If value causes.

Applies to
Aquilion 16 (TSX-101A) — installation volume 2C201-160, Appendix 28.4, "If Adjustment" (p.184-187), subsection 28.4.1 "Remote If adjustment." This is the full walkthrough that Section 15.2, "If adjustment — pointer" (p.126), sends the commissioning engineer to by name; it runs during initial system installation/commissioning, in the same slot where the identical procedure also lives in the X-ray & HV service volume (see Related).
Personnel
Intermediate — the same screen-driven skill as the X-ray & HV volume's remote If-adjustment procedure (Related): operating the If Adjustment / Exposure Test / Parameter adjustment screens, selecting the tube-voltage/focus-size/mA combinations to adjust, running Auto Exposure, and recognizing and recovering from an Over If error (Error Clear -> Details -> corrected IF value, the raise/lower direction judged from the If waveform via Signal Show) versus any other error (Error Clear -> Start, no detour). During commissioning this normally follows the X-ray-tube warm-up/rotor check (source §15.1) and precedes the tube-voltage/tube-current oscilloscope check (source §15.4). (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
11 steps
Source
2C201-160E*F p.184-187

Tools & parts

ItemPart numberQty
Personal computer terminal running the XC maintenance program (Ver. 3.0 or later)n/a in source

Full step-by-step for X-ray tube If adjustment (remote)

All 11 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related X-ray & HV guides

View all Aquilion 16 (TSX-101A) guides

Source: 2C201-160E*F p.184-187. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.