Integrated test — scan operation and image quality
Before the Aquilion 16 can be signed over, run every scan mode the console offers and prove the reconstructed images against the manufacturer's own image-noise, uniformity and CT-number specifications — the last functional gate of the installation sequence, immediately before predelivery cleaning and QC-sheet sign-off.
Applies to
Aquilion 16 (TSX-101A) — installation volume 2C201-160, Section 26, "Integrated Test" (p.160-161). Follows Section 25's DICOM environment setup, imager-preset configuration and filming operational check (p.141-159) and precedes Section 27's couch-top-end/patient-fastener fitting, unit cleaning and installation-quality-check-sheet sign-off (p.162-165).
Personnel
Intermediate — operate the console to run each scan type (scanogram, S&V, S&S, Auto-Index, dynamic, helical) and drive the standard ROI-calculation tool on the reconstructed images (250×250 pixel ROI for the phantom image-noise/CT-number-in-water table, 25×25 pixel ROI at six marked points for the TOS-phantom check); run the dedicated Q/A software for the Q/A-phantom's image-noise module; and read each measured value against the standard range printed in the tables below. (Skill assessment derived from the section's content — not an OEM-stated skill level.)
Steps
6 steps
Source
2C201-160E*F p.160-161
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| φ400 polypropylene phantom | n/a in source | |
| φ320 water phantom | n/a in source | |
| φ240 water phantom | n/a in source | |
| φ180 water phantom | n/a in source | |
| Q/A phantom + Q/A software | n/a in source | |
| TOS phantom, 320 mm dia. | n/a in source | |
| ROI-calculation tool (console software) | n/a in source |
Full step-by-step for Integrated test — scan operation and image quality
All 6 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
7-day free trial · no credit card required
Related Installation guides
Bring the patient couch and console into position9 stepsChecks before turning on the power7 stepsBasic disassembly and carrying-out of the system13 stepsFinal gantry-couch relationship check, system test and calibration data14 stepsPre-delivery: mount couch accessories, clean and fill the QC sheet10 stepsRun the console malfunction-diagnosis program (hwtest)0 steps
View all Aquilion 16 (TSX-101A) guidesRelated error codes
00000001RECON unit ended abnormally, reconstruction queue full00000002RECON unit timed out — restart the reconstruction queue00000003RECON unit failed to start — delete the reconstruction queue00000004Helical scan aborted, reconstruction request not accepted00000005RECON unit error — incorrect parameter set / Stack Scan aborted00000006AutoImageTransfer error / reconstruction sequence error
Source: 2C201-160E*F p.160-161. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.