Basic disassembly and carrying-out of the system
Before a CT system can be taken out of a room, run it through this closing sequence — erase the patient data, put the gantry and couch into a locked, de-anchored, de-energized state, and only then carry it out — exactly as laid out in Section 29 ("Basic Procedures for Disassembly and Carrying-Out of the System and Related Precautions"), the last section of the Installation volume.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Metal fittings for preventing rotation or tilting | n/a in source | n/a in source |
| Special metal fittings (eyebolts, etc.) | n/a in source | n/a in source |
| Tester | n/a in source | n/a in source |
Safety
Hazard: Contact with spilled blood or contrast medium while disassembling and carrying out the system.
Consequence: Spilled blood or contrast medium may be a source of infection.
Avoidance: Wear hospital gloves and perform appropriate disinfection before starting work. In addition, perform thorough cleaning after the CT system has been carried out.
Hazard: Performing system operations such as gantry rotation, gantry tilting, and couch vertical movement (to reach the carrying-out status) without first checking the surrounding area.
Consequence: Failure to check the surrounding area first may result in your becoming caught.
Avoidance: Before performing system operations such as gantry rotation, gantry tilting, and couch vertical movement, be sure to check the surrounding area to ensure safety.
Hazard: Disconnecting the power cables from the power distribution board without first confirming, with a tester, that no power is supplied.
Consequence: Failure to do so may result in an electric shock.
Avoidance: Before disconnecting the power cables, be sure to turn OFF the breaker at the power distribution board. In addition, use a tester to confirm that power is not supplied.
Hazard: Transporting the gantry without the tilting and rotation lock mechanisms securely engaged.
Consequence: The system may fall due to unexpected gantry tilting or rotation, resulting in personal injury.
Avoidance: Before transporting the gantry, be sure to securely engage the tilting and rotation lock mechanisms.
Hazard: Carrying out the gantry and patient couch without reviewing the installation manual's bringing-in precautions beforehand, without pre-operation checks of the tools/instruments used (wire cables, jigs, jacks, etc.), or along a carrying-out route without sufficient clearance.
Consequence: The use of inappropriate tools or instruments may allow the system to fall, resulting in personal injury. A worker may become caught between the system and a wall, etc., during carrying-out, or have hands or fingers caught and injured.
Avoidance: Carefully review in advance the precautions described in the sections related to bringing-in in the installation manual and observe them when carrying out the system. Make arrangements with a transportation agency for transport of the gantry and patient couch. Be sure to perform pre-operation checks of the tools and instruments (wire cables, jigs, jacks, etc.) to be used for carrying-out. Ensure that the carrying-out route has sufficient clearance, and do not place your hands, fingers, etc. where they may become caught and injured.
Full step-by-step for Basic disassembly and carrying-out of the system
All 13 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Installation guides
Source: 2C201-160E*F p.215-216. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.