Replace a DAS PWB (Converter-16 / Control-16)
Remove and remount a Converter-16 or Control-16 PWB by de-energizing the DAS, opening the gantry to reach the DAS chassis, swapping the board on an antistatic mat without touching its pattern surface, and reversing the same steps to close everything back up.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Standard tool set | n/a in source | 1 set |
Safety
Hazard: Opening the gantry front cover and beginning DAS PWB service without first turning OFF the circuit protector switch
Consequence: The DAS remains ON (energized) — source does not state a further consequence beyond this
Avoidance: Be sure to turn OFF the circuit protector switch before working on the DAS PWB (source labels this a CAUTION at the head of the procedure; FSELIB classifies it NOTICE per the platform's Z535 mapping, since the only stated consequence is the equipment staying energized, not personal injury)
Hazard: Touching the parts of the DAS PWB, or its pattern surface, with your hands or any other object while swapping it
Consequence: If soiled with oil, water, etc., the SD (signal/image noise) may be made worse
Avoidance: Be careful not to touch the parts of the DAS PWB or the pattern surface with your hands or other objects (source labels this a CAUTION; FSELIB classifies it NOTICE, since the only stated consequence is degraded image data, not personal injury)
Full step-by-step for Replace a DAS PWB (Converter-16 / Control-16)
All 7 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated DAS guides
Source: 2D201-065E*D p.87-88. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.