Acquire calibration data (NRA / air calibration)
Open the Aquilion 16's calibration data acquisition function from the console scan monitor — a three-click menu path that hands off to the full NRA/air-calibration workflow documented in the software volume — after waiting out the 3-hour detector warm-up that gates it.
Safety
Hazard: Starting calibration data acquisition before the main detector's temperature has stabilized after system power-on.
Consequence: Detector temperature stabilization normally takes 2 to 3 hours (varies with room temperature); calibration data acquired before it stabilizes is not reliable.
Avoidance: Wait at least 3 hours after system power is turned ON before starting calibration data acquisition.
Full step-by-step for Acquire calibration data (NRA / air calibration)
All 2 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated DAS guides
Source: 2D201-065E*D p.96. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.