ProcedureCouch/ControlAdjust

Select footswitch mode (CBTB-016A/016B)

Set the patient couch's footswitch control mode — UP/DOWN or AUTO-SET/AUTO-HOME — for a CBTB-016A or CBTB-016B couch, by entering the `fsw` command at the KGTSM boot console over a HyperTerminal session.

Applies to
Aquilion 16 (TSX-101A) — gantry/couch control volume 2D201-063, section 8.5; applies only to the patient couch CBTB-016A or CBTB-016B (source's own section title restricts this procedure to those two couch variants).
Personnel
1 technician · Basic — the whole task is a HyperTerminal session against the KGTSM boot console: connect the service cable, boot the KGTSM, type `fsw` at the `[GTS]` prompt, and enter `1` or `2` for the desired mode. No electrical measurement, cover removal, or mechanical adjustment is described, and — unlike the couch absolute-value `slide` menu in the immediately preceding subsection — this section shows no "are you sure? y/n" confirmation step: picking the wrong number sets the mode directly, so read the redisplayed "Foot sw =" line on the final summary screen carefully (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
5 steps
Source
2D201-063E*G p.161

Tools & parts

ItemPart numberQty
PCn/a in source
Service cable (GR-1)GR-1 (standard accessory)
Terminal program (e.g. Hyper Terminal)n/a in source

Full step-by-step for Select footswitch mode (CBTB-016A/016B)

All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Control guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-063E*G p.161. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.