Set couch autoslide movement range and zero-clear time
Set two related KGTSM console values for the Aquilion 16 patient couch — the couch autoslide movement range and the couch-sliding zero-clear time — over the same HyperTerminal session against the KGTSM boot console.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| PC | n/a in source | |
| Service cable (GR-1) | GR-1 (standard accessory) | |
| Terminal program (e.g. Hyper Terminal) | n/a in source |
Safety
Hazard: Leaving the couch autoslide movement range at a value other than 310 mm on a gantry unit that does not have the autoslide function.
Consequence: Not explicitly stated in source beyond the instruction itself.
Avoidance: Even for units without the autoslide function, set the movement range at 310 mm (source callout, p.154, printed as the (2) Caution under 8.1 Description).
Full step-by-step for Set couch autoslide movement range and zero-clear time
All 4 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Control guides
Source: 2D201-063E*G p.154-156. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.