Pular para o conteúdo principal
ProcedureCouch/ControlAdjust

Set couch autoslide movement range and zero-clear time

Set two related KGTSM console values for the Aquilion 16 patient couch — the couch autoslide movement range and the couch-sliding zero-clear time — over the same HyperTerminal session against the KGTSM boot console.

Applies to
Aquilion 16 (TSX-101A) — gantry/couch control volume 2D201-063, sections 8.1 ("Movement Range Setting Method for Couch Autoslide") and 8.2 ("Setting the Zero-Clear Time for Couch Sliding"); per source, the 310 mm movement-range setting in 8.1 applies even to units without the autoslide function
Personnel
1 technician · Basic-to-intermediate — both subsections are HyperTerminal sessions against the KGTSM boot console: connect the service cable to CN404, enter a numeric value at a prompt, and confirm it at a y/n readback. No electrical measurement, cover removal, or mechanical adjustment is described in this section. The KGTSM keeps whatever value is confirmed until it is rewritten, so read the confirmation line back carefully before accepting (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
4 steps
Source
2D201-063E*G p.154-156

Tools & parts

ItemPart numberQty
PCn/a in source
Service cable (GR-1)GR-1 (standard accessory)
Terminal program (e.g. Hyper Terminal)n/a in source

Safety

CAUTION

Hazard: Leaving the couch autoslide movement range at a value other than 310 mm on a gantry unit that does not have the autoslide function.

Consequence: Not explicitly stated in source beyond the instruction itself.

Avoidance: Even for units without the autoslide function, set the movement range at 310 mm (source callout, p.154, printed as the (2) Caution under 8.1 Description).

Full step-by-step for Set couch autoslide movement range and zero-clear time

All 4 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Control guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-063E*G p.154-156. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.