Adjust couch height and projector angle displays
Calibrate the operator panel's couch height readout against two named stop positions, and the operator panel's external projector angle readout against 0° and ±30°, by toggling DSW4 bit 4 on the KGTSM PWB and turning the matching VR pot at each reference point.
Applies to
Aquilion 16 (TSX-101A) — gantry/couch control volume 2D201-063, sections 8.7 (Adjustment of Couch Height Displays) and 8.8 (Adjustment of Projector Angle Display)
Personnel
1 technician · Basic-to-intermediate — the whole job is toggling one DIP-switch bit on the KGTSM PWB, driving the couch (or the external projector) to a named reference position, and turning a calibration VR pot until the operator-panel readout matches the reference value. No cover removal, electrical measurement, or component replacement is described in either section (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
10 steps
Source
2D201-063E*G p.163-164
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Screwdriver for adjustment | n/a in source |
Full step-by-step for Adjust couch height and projector angle displays
All 10 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Control guides
Set couch autoslide movement range and zero-clear time4 stepsPatient couch control troubleshooting (CBTB-016A/016B)11 stepsSet couch horizontal movement absolute-value control data9 stepsCouch slide servo amplifier operation and maintenance tool (CBTB-016A/016B)10 stepsSet couch-top sliding fine-adjustment function5 stepsCheck the couch-top speed2 steps
View all Aquilion 16 (TSX-101A) guidesSource: 2D201-063E*G p.163-164. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.