ProcedureCouch/ControlAdjust

Set couch horizontal movement absolute-value control data

Set the four sensor-to-sensor distances the KGTSM uses to control where the patient couch top's horizontal (IN/OUT) travel starts and stops, by walking a HyperTerminal session through the KGTSM boot console's `slide` menu and picking (or hand-entering) the distance set that matches the couch actually installed.

Applies to
Aquilion 16 (TSX-101A) — gantry/couch control volume 2D201-063, section 8.3; covers couch variants CBTB-014A, CBTB-016A, CBTB-016B, Selfpropelled, and any other couch via the custom-entry path
Personnel
1 technician · Basic-to-intermediate — the whole job is a HyperTerminal session against the KGTSM boot console: connecting a service cable to CN404, navigating a menu of `y`/`n` and numeric prompts, and (for a non-catalog couch) hand-entering four distance values. No electrical measurement, cover removal, or mechanical adjustment is described in this section. Picking the wrong couch-type entry, or mistyping a custom distance, sets the wrong travel-limit data for the physical sensors installed — so read back the confirmation screen carefully before accepting (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
9 steps
Source
2D201-063E*G p.157-159

Tools & parts

ItemPart numberQty
PCn/a in source
Service cable (GR-1)GR-1 (standard accessory)
Terminal program (e.g. Hyper Terminal)n/a in source

Full step-by-step for Set couch horizontal movement absolute-value control data

All 9 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Control guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-063E*G p.157-159. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.