ProcedureCouch/ControlMeasure
Check the couch-top speed
Time the patient couch top's marked IN/OUT travel with a stopwatch against the manufacturer's speed-and-tolerance tables, to confirm the couch top is actually running at its two commanded speeds — 10 mm/s and either 130 mm/s (CBTB-016A/016B) or 100 mm/s (CBTB-014A).
Applies to
Aquilion 16 (TSX-101A) — gantry/couch control volume 2D201-063, section 8.9 ("Checking the Couch Top Speed"); the 10/130 mm/s table applies to the CBTB-016A or CBTB-016B patient couch, the 10/100 mm/s table applies to the CBTB-014A patient couch. Source gives no table for a Selfpropelled couch in this section.
Personnel
2 technicians · Basic — one person runs the couch from the operating panel while the other times the marked travel with a stopwatch and reads the distance off the taped scale; no electrical measurement, cover removal, or KGTSM console session is involved (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
2 steps
Source
2D201-063E*G p.165-166
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Measuring tape | n/a in source | |
| Stopwatch | n/a in source |
Full step-by-step for Check the couch-top speed
All 2 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Control guides
Set couch autoslide movement range and zero-clear time4 stepsPatient couch control troubleshooting (CBTB-016A/016B)11 stepsAdjust couch height and projector angle displays10 stepsSet couch horizontal movement absolute-value control data9 stepsCouch slide servo amplifier operation and maintenance tool (CBTB-016A/016B)10 stepsSet couch-top sliding fine-adjustment function5 steps