BB009000 — Error in Marker Controller (LAMS device)
Canon Medical (Toshiba) CT error code, reported by the LAMS device.
What BB009000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code BB009000 indicates: Error in Marker Controller (LAMS device). The code originates in the LAMS device section of the LAMS device. Severity level A0 (Information / dialog).
Reported action: Exercise the marker function to reproduce the fault and watch whether the marker hardware responds at all — that separates a dead controller from a signalling problem. Check the Marker Controller's supply voltage, its cabling and its connectors to the hardware it drives, then repeat the operation; if the code returns, follow the LAMS troubleshooting flow in the service manual..
Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems. The code is raised by the LAMS device and reports a fault in the Marker Controller it supervises.
Full troubleshooting for BB009000
The complete diagnosis, cause tree, and step-by-step remedy for the LAMS device subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated LAMS device codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.