Pular para o conteúdo principal
Information / dialogLAMS device

BB009000Error in Marker Controller (LAMS device)

Canon Medical (Toshiba) CT error code, reported by the LAMS device.

What BB009000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code BB009000 indicates: Error in Marker Controller (LAMS device). The code originates in the LAMS device section of the LAMS device. Severity level A0 (Information / dialog).

Reported action: Exercise the marker function to reproduce the fault and watch whether the marker hardware responds at all — that separates a dead controller from a signalling problem. Check the Marker Controller's supply voltage, its cabling and its connectors to the hardware it drives, then repeat the operation; if the code returns, follow the LAMS troubleshooting flow in the service manual..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems. The code is raised by the LAMS device and reports a fault in the Marker Controller it supervises.

Full troubleshooting for BB009000

The complete diagnosis, cause tree, and step-by-step remedy for the LAMS device subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related LAMS device codes

Related service procedures

Discuss BB009000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.