Pular para o conteúdo principal
ProcedureConsole (Recon)Replacement/FRUReplace

Remove and install an RDD fan unit (Rorke Data)

Remove and reinstall either of the two independently replaceable fan units in the Rorke Data-manufactured raw data disk array (RDD/JBOD) chassis — each secured by three screws — without disturbing the other fan.

Applies to
Aquilion 16 (TSX-101A) — Console (Recon) volume, source 2D201-079, §6.7.3.5 ("Procedures for removing/installing the fan unit"), p.89. This procedure is specific to the raw-data-disk-array (RDD/JBOD) chassis manufactured by **Rorke Data** (§6.7.3, the second of the two JBOD vendors documented in the RDD chapter). Source gives no equivalent fan-unit procedure for the Medea-manufactured JBOD documented earlier in the same chapter (§6.7.2, which ends at the disk-unit removal procedure, §6.7.2.4) — see toshiba-ct-aquilion16-console-rdd-jbod-medea, Related.
Personnel
1 technician · Basic — remove the three screws securing the fan unit being serviced, pull it out, then reinstall by aligning the fan unit's power connector with the matching power connector in the cabinet and refitting the three screws. No meter, test equipment, fastener size/thread designator, or torque value is stated anywhere in this section (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
5 steps
Source
2D201-079E*H p.89

Tools & parts

ItemPart numberQty
FAN1 securing screwsn/a in source3
FAN2 securing screwsn/a in source3
Replacement fan unit (FAN1 or FAN2)n/a in source1

Full step-by-step for Remove and install an RDD fan unit (Rorke Data)

All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Console (Recon) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-079E*H p.89. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.