BB008000 — Error in LMHC (LAMS device)
Canon Medical (Toshiba) CT error code, reported by the LAMS device.
What BB008000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code BB008000 indicates: Error in LMHC (LAMS device). The code originates in the LAMS device section of the LAMS device. Severity level A0 (Information / dialog).
Reported action: Read the full LAMS entry in the service log to see when LMHC failed and whether other LAMS units reported at the same time. Then check the LMHC unit's supply voltage and its cabling and connectors inside the LAMS assembly, re-seat them and repeat the operation that raised the code; if LMHC still faults, follow the LAMS troubleshooting flow in the service manual before ordering parts..
Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems. The code is raised by the LAMS device and reports a fault in LMHC, one of the units the LAMS device supervises.
Full troubleshooting for BB008000
The complete diagnosis, cause tree, and step-by-step remedy for the LAMS device subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated LAMS device codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.