000001F8 — Database full — image or raw data storage
Canon Medical (Toshiba) CT error code, reported by the FSYS / RSYS (exam database).
What 000001F8 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 000001F8 indicates: Database full — image or raw data storage. The code originates in the FSYS / RSYS (exam database) section of the FSYS / RSYS (exam database). Severity level 1 (Information / dialog).
Reported action: The image database (FSYS) or the raw data database (RSYS) is full. Archive or delete older studies to free space, then retry storing or reconstructing..
Toshiba/Canon CT consoles on Aquilion RX/V4-62R004, V3.20R003, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and Aquilion 16. The OEM text reads "Database full ( Image )." in the FSYS table on all six platform columns, and "Database full ( Raw Data )." in the RSYS table on Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2.
Full troubleshooting for 000001F8
The complete diagnosis, cause tree, and step-by-step remedy for the FSYS / RSYS (exam database) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated FSYS / RSYS (exam database) codes
Related service procedures
Source: errmsg.eng — 60:Device FSYS / 70:Device RSYS tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.