Pular para o conteúdo principal
Information / dialogFSYS / RSYS (exam database)

000001F7Exceeded the maximum number of exams the database can hold

Canon Medical (Toshiba) CT error code, reported by the FSYS / RSYS (exam database).

What 000001F7 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 000001F7 indicates: Exceeded the maximum number of exams the database can hold. The code originates in the FSYS / RSYS (exam database) section of the FSYS / RSYS (exam database). Severity level 1 (Information / dialog).

Reported action: The image database (FSYS) or the raw data database (RSYS) has reached the maximum number of exams it can store. Archive or delete older exams to free entries before scanning again..

Toshiba/Canon CT consoles on Aquilion RX/V4-62R004, V3.20R003, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and Aquilion 16. The OEM text reads "Exceeded max. exams in Database." in the FSYS table on all six platform columns, and "Exceeded max. exams in Database (Raw Data)." in the RSYS table on Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2.

Full troubleshooting for 000001F7

The complete diagnosis, cause tree, and step-by-step remedy for the FSYS / RSYS (exam database) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related FSYS / RSYS (exam database) codes

Related service procedures

Discuss 000001F7 with other field engineers in the FSELIB forum

Source: errmsg.eng — 60:Device FSYS / 70:Device RSYS tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.