000001D2 — Exceeded the maximum images or raw data sets allowed in one exam
Canon Medical (Toshiba) CT error code, reported by the FSYS / RSYS (exam database).
What 000001D2 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 000001D2 indicates: Exceeded the maximum images or raw data sets allowed in one exam. The code originates in the FSYS / RSYS (exam database) section of the FSYS / RSYS (exam database). Severity level 1 (Information / dialog).
Reported action: The current exam has hit the system limit for images (FSYS) or raw data sets (RSYS) that a single exam can hold. Close the exam and start a new one to continue scanning; if this happens routinely, review the protocol's image and raw-data output against the per-exam limit..
Toshiba/Canon CT consoles on Aquilion RX/V4-62R004, V3.20R003, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and Aquilion 16. The OEM text reads "Exceeded max. images in an exam." in the FSYS table on all six platform columns, and "Exceeded max. raws in an exam." in the RSYS table on Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2.
Full troubleshooting for 000001D2
The complete diagnosis, cause tree, and step-by-step remedy for the FSYS / RSYS (exam database) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated FSYS / RSYS (exam database) codes
Related service procedures
Source: errmsg.eng — 60:Device FSYS / 70:Device RSYS tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.