Gantry safety-circuit operation checks via KGTSM (tilt/vertical/horizontal, runout, rotation error) — 1.38/1.39/1.40
Exercise the KGTSM PWB's dedicated safety circuit end-to-end over a HyperTerminal session — the tilt, couch-vertical, and couch-horizontal double safety signals and their timers (item 1.38), the couch horizontal-movement runout lock (item 1.39), and, on TSX-101A/5, the rotation-ERROR detection circuit (item 1.40) — confirming each one still shuts the gantry down exactly the way it is supposed to.
Safety
Hazard: not performing item (c) — disconnecting CN451 of the KGTSM PWB — before checking a safety circuit that turns out to be defective
Consequence: the gantry may move unexpectedly, resulting in injury
Avoidance: be sure to perform item (c) exactly as given: disconnect CN451 of the KGTSM PWB before proceeding to the safety-circuit checks below
Hazard: operating the gantry/couch from the console keyboard while this KGTSM safety-circuit check is running
Consequence: not spelled out beyond the imperative in this source passage — discontinuing console operation is stated as a flat requirement for the duration of the check, not qualified with its own consequence text here
Avoidance: turn OFF the gantry/couch operation switch on the console keyboard before starting (step (e)) and keep console operation discontinued for the whole check
Full step-by-step for Gantry safety-circuit operation checks via KGTSM (tilt/vertical/horizontal, runout, rotation error) — 1.38/1.39/1.40
All 13 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry guides
Source: 2Y201-061E*E p.77-85. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.