Operational checks of gantry, couch, detector and X-ray tube
Confirm the gantry, patient couch, detector and X-ray tube all operate to standard — working through the operating panel, the KGTSM maintenance panel, the footswitch, an optical data-transmission check, an emergency couch-release check, a detector-temperature check, and an X-ray tube warm-up/seasoning/voltage-current check — before moving on to the phantom and test-scan checks that follow.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Storage oscilloscope | n/a in source | |
| PC running the XC maintenance program | n/a in source | |
| Cable | n/a in source | |
| Thermoregulator (replacement) | n/a in source |
Safety
Hazard: Performing the X-ray exposure needed for the tube-voltage/tube-current check (Step 10 below) while a worker remains inside the CT room
Consequence: Unnecessary X-ray exposure to that worker (source states no further detail on dose or injury)
Avoidance: After setting up the storage oscilloscope, exit the CT room and confirm that no person is in the CT room before performing the exposure
Full step-by-step for Operational checks of gantry, couch, detector and X-ray tube
All 18 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry guides
Source: 2C201-160E*F p.122-127. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.